Gas Configuration
The pressure, flow rate, and quantity configuration of the various gases in the clean operating department are shown in Tables 8.2 and 8.3.
Table 8.2 Minimum number of connection points configured for the main gas types in the clean operating department
| Department | Oxygen | Negative Pressure Suction | Compressed Air | Nitrous Oxide | Nitrogen |
|---|---|---|---|---|---|
| Operating Room | 2 | 2 | 1 | 1 | 1 |
| Recovery Room | 1 | 2 | 1 | ||
| Pre-anesthesia Room | 1 | 1 | 1 | 1 |
Table 8.3 Terminal Pressure, Flow Rate, and Daily Usage Time
| Gas Type | Single Outlet Pressure/MPa | Flow Rate | Flow Rate | Flow Rate |
|---|---|---|---|---|
| L/min | Daily Usage Time/min | Simultaneous Usage Rate/% | ||
| Oxygen | 0.4–0.45 | 10–80 | 120 (Recovery Room 144) | 50–100 |
| Negative Pressure Suction | -0.035–-0.075 | 30 | 120 (Recovery Room 144) | 100 |
| Compressed Air | 0.45–0.9 | 60 | 60 | 80 |
| Nitrogen | 0.8–0.9 | 230 | 30 | 10–60 |
| Nitrous Oxide | 0.4–0.5 | 4 | 120 | 50–100 |
| Argon | 0.35–0.4 | 0.5–15 | 120 | 80 |
| Carbon Dioxide | 0.35–0.4 | 10 | 60 | 30 |
Compressed air should, wherever possible, be installed with two types of outlets for two different pressure requirements: one outlet has an output pressure of 0.4 MPa and is mainly used for ventilator power, spray therapy, and mixing with other types of gases, where the pressure used should not be too high; the other outlet has an output pressure of 0.9 MPa, which is relatively high, and is used for high-speed power tools such as pneumatic drills and pneumatic saws, mainly for open-chest and cranial surgery, and therefore requires a relatively high pressure.