Film Viewer
Embedded X-ray film viewers are available in several sizes, which can be selected according to the complexity of the surgery. In general, complex surgeries are often at high
Figure 6.15 X-ray Film Viewer performed in the higher-grade operating room, and the film viewer selected is also of larger specifications. However, in lower-grade operating rooms, at least a triple-type film viewer should be selected. Its installation position is preferably on the wall opposite the chief surgeon. The appearance of the film viewer is shown in Figure 6.15.
The specifications of film viewers are given in Table 6.3. Abroad, there are also light-shielding film viewers that expose only a single particular film, and there are imitated products in China as well. The dimensions of film viewers vary among manufacturers, so the opening should be cut according to the actual sample during installation. Film viewer products must be provided with a safety grounding terminal, which shall be grounded at installation.
Film viewers may be configured with reference to the specifications and dimensions in Table 6.3
Table 6.3 Examples of Film Viewer Specifications
| Number of films (large size) | Number of films (small size) | Type | Dimensions/mm W(Length) | Dimensions/mm H(Height) | Dimensions/mm D(Thickness) | Number of fluorescent tubes 15W/tube | Current 220V | Remarks |
|---|---|---|---|---|---|---|---|---|
| 2 films | 3 films | MS.JU215 | 1 000 | 600 | 120 | 7 | 0.5 | |
| 3 films | 4 films | MS.JU315 | 1 200 | 600 | 120 | 9 | 0.7 | |
| 4 films | 5 films | MS.JU415 | 1 600 | 600 | 120 | 12 | 0.9 | |
| 5 films | 6 films | MS.JU515 | 2 000 | 600 | 120 | 15 | 1.1 | |
| 6 films | 7 films | MS.JU615 | 2 400 | 600 | 120 | 18 | 1.3 | 3 films, 2 columns |
| 4 films | 6 films | MS.JU425 | 1 000 | 1 000 | 120 | 14 | 1.0 | |
| 6 films | 8 films | MS.JU625 | 1 200 | 1 000 | 120 | 18 | 1.3 | |
| 8 films | 10 films | MS.JU825 | 1 600 | 1 000 | 120 | 24 | 1.7 | 3 films, 2 rows, 2 columns |
| 10 films | 12 films | MS.JU1025 | 2 000 | 1 000 | 120 | 30 | 2.1 | |
| 12 films | 14 films | MS.JU1225 | 2 400 | 1 000 | 120 | 36 | 2.5 | |
| Film viewer for plastic surgery | Film viewer for plastic surgery | MS.JU825 | 2 400 | 1 000 | 120 | 15W×2 tubes 20W×7 tubes | 2.5 |